Does intra-osseous fixation with the IO FiX improve force and contact area in foot and ankle fusions?

L. Parker, P. Ray, S. Grechenig, W. Grechenig

1 The Royal National Orthopaedic Hospital, Middlesex, United Kingdom 

2 Barnet Hospital, London, United Kingdom

3 Medical University of Graz, Department of Traumatology, Graz, Austria

Introduction: When inserting a lag-screw across an arthrodesis, stress is concentrated under the screw head risking asymmetrical force distribution and fracture of the cortical bone bridge. The IO FiX (Extremity Medical, NJ USA) is a new intraosseous device comprising an X-Post on one side of and parallel to the arthrodesis and a lag-screw inserted through the head of the X-Post which reinforces the cortical bone bridge. The X-Post behaves as an internal washer improving force distribution across the arthrodesis. Being intraosseous, near to the neutral axis of bend also means the device is fatigue-resistant and soft tissue irritation is reduced.

Methods: The IO FiX has not been independently verified and therefore we analysed its performance in a human cadaveric ankle model. Our null hypothesis was there is no difference in force generation and contact area in an ankle arthrodesis when the IO FiX is compared with partially-threaded lag-screws. We used ten randomized cadaver ankles with a mean age of seventy-one years (44-84 years) prepared with flat arthrodesis cuts. A Tek-scan (Boston, USA) pressure transducer was used to measure force and contact area produced when the IO FiX was compared with a standard lag-screw and washer.

Results: The median average force in the IO FiX group was 3.95 kg and 2.35 kg in the lag-screw group (p=< 0.01 Wilcoxon signed-rank). The IO FiX was able to create a more uniform contact area within the arthrodesis with a median average of 3.41 cm2 compared with 2.42 cm2 in the lag-screw group (p=< 0.03 Wilcoxon signed rank).

Conclusion: Our results suggest the IO FiX improves force generation and contact area across the arthrodesis. With the theoretical advantages of reduced soft tissue irritation and a lower risk of fatigue failure, the IO FiX offers a significant advantage compared with traditional fixation techniques.

Previous Article Does functional outcome depend on the quality of the fracture fixation? Inter-observer variability of radiological evaluation of surgical ankle fracture fixation. Analysis of 61 cases
Next Article Does the presence of intra-articular pathology affect the outcome following modified Brostrom repair for lateral ligament instability of the ankle?
Print


Click thumbnail below to view poster / thumbnail:

Archive of Abstracts

2024   -   Prize Winners  |  All Abstracts
2023   -   Prize Winners  |  All Abstracts
2022   -   Prize Winners  |  All Abstracts
2021   -   Prize Winners  
2019   -   Podium  |  Poster
2018   -   Podium  |  Poster
2017   -   Podium  |  Poster
2016   -   Podium  |  Poster
2015   -   Podium  |  Poster
2014   -   Podium  |  Poster
2013   -   Podium  |  Poster
2011   -   All Abstracts
2009   -   All Abstracts
2008   -   All Abstracts
2007   -   All Abstracts
2006   -   All Abstracts
2005   -   All Abstracts
2004   -   All Abstracts
2002   -   All Abstracts
2001   -   All Abstracts
2000   -   All Abstracts
1999   -   All Abstracts
1998   -   All Abstracts
1997   -   All Abstracts
1996   -   All Abstracts
1995   -   All Abstracts
1994   -   All Abstracts
1993   -   All Abstracts
1991   -   All Abstracts
1990   -   All Abstracts
1989   -   All Abstracts
1987   -   All Abstracts
1985   -   All Abstracts
1983   -   All Abstracts